Article ID Journal Published Year Pages File Type
7231626 Biosensors and Bioelectronics 2015 40 Pages PDF
Abstract
In this work, a sensitive amperometric biosensor for hydrogen peroxide based on synergetic catalysis of hemoglobin and porous Pd@Fe3O4-MWCNT nanocomposite has been constructed. With attention to the utilities of large surface area and outstanding catalytic performance, Pd@Fe3O4-MWCNT nanocomposite was employed as the nano-stabilizer for the immobilization of hemoglobin (Hb). The immobilized Hb on the surface of nanocomposite as an electrochemical biosensor efficiently catalyzed the reduction of hydrogen peroxide, amplified the electrochemical signal and enhanced the sensitivity. Results of voltammetry and electrochemical impedance examinations showed that the nanocomposite could enhance the electron conductivity and provide more sites for the immobilization of Hb. A linear response from 0.2-500 µM with detection limit of 0.063 µM for hydrogen peroxide was achieved. The apparent Michaelis-Menten constant KappM value was 21 µM. Thus, the nanocomposite could be applied for fabrication of a third generation biosensor for hydrogen peroxide with high sensitivity, selectivity and low detection limit. The excellent performance of the biosensor indicated its promising prospect as a valuable tool in simple and fast hydrogen peroxide detection in environmental and clinical applications.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,