2-Aminobenzyl amine (2-ABA) functionalized graphene is proposed for the ultrasensitive immunosensing of Cardiac Troponin I (cTnI). 2-ABA was electrochemically polymerized on the graphene decorated interdigitated electrode to obtain the amine functionalized graphene (f-GN). The f-GN electrode was then modified with monoclonal anti-cTnI antibodies via Schiff reaction based chemistry. Detailed characteristics of the processes involved and the finally developed antibody conjugated f-GN interdigitated electrode have been studied. The above micro-device was used in a drain source configuration for the sensing of cTnI. A wide dynamic linear range of antigen detection (0.01-1Â ng/mL) is achieved with the limit of detection of 0.01Â ng/mL. The utility of the proposed sensing technique is demonstrated by successfully testing the antigen concentration in spiked serum samples.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Satish K. Tuteja, Manil Kukkar, C.R. Suri, A.K. Paul, Akash Deep,