Abstract
The streaking mode cytometer has been used for the analysis of SYTO-9 labeled THP-1 human monocytes in buffer and in blood. Samples of cells at 1 cell/mL and 0.1 cell/mL were analyzed in 30 mL with flow rates set to 10 mL/min and frame rates of 4 fps (frame per second). For the target of 1 cell/mL, an average concentration of 0.91 cell/mL was measured by cytometry, with a standard error of 0.03 (C95=0.85-0.97). For the target of 0.1 cell/mL, an average concentration of 0.083 cell/mL was measured, with a standard error of 0.01 (C95=0.065-0.102). Whole blood was also spiked with SYTO-9 labeled cells to a concentration of 10 cell/mL, and the average flow cytometry measurement was 8.7 cells/mL (i.e. 0.87 cells/mL in diluted blood) with a 95% CL of 8.1-9.2 cells/mL. This demonstrated the ability to detect rare cells in blood with high accuracy. Such detection approaches for rare cells have many potential clinical applications. Furthermore, the simplicity and low cost of this device may enable expansion of cell-based clinical diagnostics, especially in resource-poor settings.