Abstract
In this paper, a pen-on-paper electrochemiluminescence (PoP-ECL) device was entirely hand drawn and written in commercially available crayon and pencil in turn for the first time, and a constant potential-triggered sandwich-type immunosensor was introduced into the PoP-ECL device to form a low-cost ECL immunodevice proof. Each PoP-ECL device contained a hydrophilic paper channel and two PoP electrodes, and the PoP-ECL device was produced as follows: crayon was firstly used to draw hydrophobic regions on pure cellulose paper to create the hydrophilic paper channels followed with a baking treatment, and then a 6B-type black pencil with low resistivity was applied for precision writing, as the PoP electrodes, across the hydrophilic paper channel. For further point-of-care testing, a portable, low-cost rechargeable battery was employed as the power source to provide constant potential to the PoP electrodes to trigger the ECL. Using Carbohydrate antigen 199 as model analyte, this PoP-ECL immunodevice showed a good linear response range from 0.01-200Â UÂ mLâ1 with a detection limit of 0.0055Â UÂ mLâ1, a high sensitivity and stability. The proposed PoP-ECL immunodevice could be used in point-of-care testing of other tumor markers for remote regions and developing countries.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Hongmei Yang, Qingkun Kong, Shaowei Wang, Jinmeng Xu, Zhaoquan Bian, Xiaoxiao Zheng, Chao Ma, Shenguang Ge, Jinghua Yu,