Abstract
In the present work, a novel strategy based on overoxidized polyimidazole (PImox) and graphene oxide (GO) copolymer modified electrode was proposed for the simultaneous determination of ascorbic acid (AA), dopamine (DA), uric acid (UA), guanine (G) and adenine (A). The copolymer was characterized by the scanning electron microscopy (SEM), atomic force microscopy (AFM), Fourier transform infrared (FT-IR), X-ray photoelectron spectroscopy (XPS) and electrochemical impedance spectroscopy (EIS). Due to the synergistic effects between PImox and GO, the proposed electrode exhibited excellent electrochemical catalytic activities and high selectivity and sensitivity toward the oxidation of AA, DA, UA, G and A. The peak separations between AA and DA, AA and UA, UA and G, and G and A were 140 mV, 200 mV, 380 mV and 300 mV, respectively. The linear response ranges for AA, DA, UA, G and A were 75-2275 μM, 12-278 μM, 3.6-249.6 μM, 3.3-103.3 μM and 9.6-215 μM, respectively, and corresponding detection limits were 18 μM, 0.63 μM, 0.59 μM, 0.48 μM and 1.28 μM.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Xiaofang Liu, Ling Zhang, Shaping Wei, Shihong Chen, Xin Ou, Qiyi Lu,