Abstract
This report describes the design, development, validation and long-term performance of tPA clot lysis activity assay using Advanced Chemistry Line Total Operational Performance (ACL TOP)™ Homeostasis Testing System. The results of the study demonstrated robust and stable performance of the analytical method. The accuracy of the assay, expressed by percent recovery is 98-99%. The intermediate precision and repeatability precision, expressed as Relative Standard Deviation (RSD), was 3% and less than 2% respectively. The validated range is from 70% to 130% of the target potency of 5.8 × 105 IU/mg. The linearity of this range, expressed in correlation coefficient, is 0.997. After the assay is transferred to a QC laboratory, the assay retained high accuracy and precision with a success rate of >99%.