Left ventricular assist devices are associated with thromboembolic events, which are potentially caused by altered intraventricular flow. Due to patient variability, differences in apical wall thickness affects cannula insertion lengths, potentially promoting unfavourable intraventricular flow patterns which are thought to be correlated to the risk of thrombosis. This study aimed to present a 3D multiscale computational fluid dynamic model of the left ventricle (LV) developed using a commercial software, Ansys, and evaluate the risk of thrombosis with varying inflow cannula insertion lengths in a severely dilated LV. Based on a HeartWare HVAD inflow cannula, insertion lengths of 5, 19, 24 and 50â¯mm represented cases of apical hypertrophy, typical ranges of apical thicknesses and an experimental length, respectively. The risk of thrombosis was evaluated based on blood washout, residence time, instantaneous blood stagnation and a pulsatility index. By introducing fresh blood to displace pre-existing blood in the LV, after 5 cardiac cycles, 46.7%, 45.7%, 45.1% and 41.8% of pre-existing blood remained for insertion lengths of 5, 19, 24 and 50â¯mm, respectively. Compared to the 50â¯mm insertion, blood residence time was at least 9%, 7% and 6% higher with the 5, 19 and 24â¯mm insertion lengths, respectively. No instantaneous stagnation at the apex was observed directly after the E-wave. Pulsatility indices adjacent to the cannula increased with shorter insertion lengths. For the specific scenario studied, a longer insertion length, relative to LV size, may be advantageous to minimise thrombosis by increasing LV washout and reducing blood residence time.
Related Topics
Physical Sciences and Engineering
Engineering
Biomedical Engineering
Authors
Sam Liao, Michael Neidlin, Zhiyong Li, Benjamin Simpson, Shaun D. Gregory,