Abstract

A portable device for calibration of trace humidity sensors and an adopted calibration procedure have been developed. The calibration device is based on humidity generation by permeating water through polymeric membrane tubes. Water vapour transmission rates for various polymers were experimentally determined in order to select the most suitable polymeric material. The developed trace humidity generator consists of a gas-flow polymeric hose immersed in a water reservoir thermostated by a sensor-controlled heater. Mole fractions of water vapour between 1 μmol mol−1 and 350 μmol mol−1 (equivalent to frost-point temperatures from −76 °C to −31 °C) were generated by varying either the operating temperature or gas flow. The operating temperature can be varied from 20 °C to 60 °C and kept stable within 0.1 K. Uncertainty analysis indicated that the trace humidity generator produces gas flows of constant humidity amounts with a relative expanded uncertainty less than 3.4% (k = 2) of the generated value.