Abstract

A simple method for reducing the uncertainty of reference standards has been proposed for gauge block comparator calibration. Errors due to the reference value of the length difference for mechanical comparator evaluation can be reduced by using the average of two reference values. Two reference values of length difference are obtained by the appropriate combination of three gauge blocks. Errors caused by reference gauges are canceled and averaged. An error reduction effect was successfully demonstrated using an actual mechanical comparator.