Article ID Journal Published Year Pages File Type
731292 Measurement 2013 5 Pages PDF
Abstract

•We identify several sources of uncertainty in quantification of linezolid.•We evaluate how each uncertainty source contributes to total uncertainty.•We establish a procedure to calculate combined and expended uncertainties.

The aim of this work was to identify the main sources of uncertainty, quantify the standard deviation of each source of uncertainty, and calculated combined and expended uncertainties for the determination of linezolid in injectable dosage forms by UV spectrophotometry. The contributions of precision, linearity and weight of linezolid reference standard are the most significant, contributing with about 77% of the overall uncertainty. The uncertainties on the absorbances (sample and standard), volumetric flasks and volumetric pipettes have virtually no influence on the overall uncertainty. The estimated uncertainty of the spectrophotometric method for determination of linezolid was adequate for the scope of the method.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide