731580 | Measurement | 2010 | 5 Pages |
Abstract
This paper describes a measurement method developed at National Institute of metrological Research (INRIM) to calibrate picoammeters in dc current from 100 pA to 100 nA. The current source is based on a traceable to the dc resistance national standard 10 × 100 MΩ Hamon resistor developed at INRIM and on a traceable to the dc voltage national standard high precision dc voltage calibrator. The expanded uncertainties of the method for the calibration of picoammeters span from 9.4 × 10−4 for the gain of a picoammeter at 100 pA to 4.0 × 10−4 for the gain at 100 nA. A detailed uncertainties budget at 10 nA level and the results of a comparison with a different technique are also reported.