Article ID Journal Published Year Pages File Type
734378 Optics & Laser Technology 2014 7 Pages PDF
Abstract

•LIBS system was applied to detect fluoride content in semi-fluid toothpaste samples.•The system was optimized to achieve optically thin and LTE plasma.•The strong atomic transition line of fluorine at 731.102 nm was used as the marker.•The system was able to detect fluoride concentration of about 1300–1750 ppm.

Fourth harmonic of a pulsed Nd:YAG laser (wavelength 266 nm) in combination with high resolution spectrograph equipped with Gated ICCD camera has been employed to design a high sensitive analytical system. This detection system is based on Laser Induced Breakdown Spectroscopy and has been tested first time for analysis of semi-fluid samples to detect fluoride content present in the commercially available toothpaste samples. The experimental parameters were optimized to achieve an optically thin and in local thermo dynamic equilibrium plasma. This improved the limits of detection of fluoride present in tooth paste samples. The strong atomic transition line of fluorine at 731.102 nm was used as the marker line to quantify the fluoride concentration levels. Our LIBS system was able to detect fluoride concentration levels in the range of 1300–1750 ppm with a detection limit of 156 ppm.

Related Topics
Physical Sciences and Engineering Engineering Electrical and Electronic Engineering
Authors
, , , , , ,