Abstract
The results of calibration experiments for individual modules suggested that pumps are yielding accurate flow rate, autosampler is aspirating the correct injection volume irrespective to position or volume, and column oven and autosampler chambers are maintaining steady temperatures. Holistic calibration verified the accuracy, precision, repeatability and reproducibility of the instrument with insignificant carryover.