Article ID Journal Published Year Pages File Type
7606215 Journal of the American Society for Mass Spectrometry 2007 9 Pages PDF
Abstract
FTMS performance parameters such as limits of detection, dynamic range, sensitivity, and even mass accuracy and resolution can be greatly improved by enhancing its detection circuit. An extended investigation of significant design considerations for optimal signal-to-noise ratio in an FTMS detection circuit are presented. A low noise amplifier for an FTMS is developed based on the discussed design rules. The amplifier has a gain of ≈3500 and a bandwidth of 10 kHz to 1 MHz corresponding to m/z range of 100 Da to 10 kDa (at 7 Tesla). The performance of the amplifier was tested on a MALDI-FTMS, and has demonstrated a 25-fold reduction in noise in a mass spectrum of C60 compared with that of a commercial amplifier.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,