Article ID Journal Published Year Pages File Type
7607902 Journal of Chromatography A 2018 34 Pages PDF
Abstract
Carbon-silica hybrid nanofibers as high performance coatings for solid-phase microextraction fibers were used for analyzing some pesticides by using gas chromatography-corona discharge ion mobility spectrometry. To that end, the fibers were prepared by carbonizing sol-gel based on electrospun polyacrylonitrile and tetraethyl orthosilicate nanofibers as carbon and silica precursors, respectively. Different parameters affecting the electrospinning and the extraction processes including spinning distance, voltage, feeding rate, stirring rate, salt concentration, temperature and extraction time were optimized by response surface methodology. The method involved deionized water samples spiked with pesticides at different concentration levels. The calibration curves were linear in the ranges of 0.1-20 and 0.05-20 μg L−1 with determination coefficients (R2) of 0.9976 and 0.9928 for malathion and chlorpyrifos, respectively. The limits of detection of 0.032 and 0.019 μg L−1 and the limits of quantification of 0.1 and 0.05 μg L−1 were found for malathion and chlorpyrifos, respectively. Acceptable reproducibility values were obtained with relative standard deviations (RSD, n = 3) lower than 6 and 15%, for intra-day and inter-day precision, respectively. Finally, the relative recoveries of the proposed method were calculated in the range of 80-111% for real samples.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,