Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7608096 | Journal of Chromatography A | 2018 | 22 Pages |
Abstract
Within this project the two most promising methods for endogenous steroid profiling were investigated and compared: the “gold standard” GC-MS and the orthogonal separation technique SFC-MS/MS. Different derivatisation procedures for gas chromatographic detection were explored and the formation of multiple derivatives described and confirmed. Taken together, none of the investigated derivatisation procedures provided acceptable results for further method development to meet the requirements of this project. SFC with its unique selectivity was able to overcome these issues and to distinguish all selected steroids, including (proâ)gestagens, androgens, corticoids, estrogens, and steroid sulfates with appropriate selectivity. Valued especially in the separation of enantiomeric analytes, SFC has shown its potential as alternative to GC. The successful separation of 51 steroids and steroid sulfates on different columns is presented to demonstrate the potential of SFC in endogenous steroid profiling.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Juliane Teubel, Bernhard Wüst, Carola G. Schipke, Oliver Peters, Maria Kristina Parr,