Article ID Journal Published Year Pages File Type
7612064 Journal of Chromatography A 2015 9 Pages PDF
Abstract
A method to analyse 2-methylpentanoic, 3-methylpentanoic and 4-methylpentanoic acids as well as cyclohexanecarboxylic acid has been developed and applied to wine and other alcoholic beverages. Selective isolation with solid phase extraction, derivatization with 2,3,4,5,6-pentafluorobenzyl bromide at room temperature for 30 minutes, and further analysis by gas chromatography-mass spectrometry in negative chemical ionization mode provides detection limits between 0.4 and 2.4 ng/L. Good linearity up to 3.6 μg/L, satisfactory reproducibility (RSD < 10%) and signal recovery of around 100% represent a robust method of analysis. Concentration data of these analytes in wine and other alcoholic beverages are reported for the first time. The levels found ranged from the method detection limits to 2630 ng/L, 2040 ng/L and 3810 ng/L for 2-, 3- and 4-methylpentanoic acids, respectively, and to 1780 ng/L for cyclohexanecarboxylic acid. There are significant differences depending on the type of wine or beverage. Distilled beverages, beer and aged wines have higher contents in methylpentanoic and cyclohexanecarboxylic acids.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry