Article ID Journal Published Year Pages File Type
7612169 Journal of Chromatography A 2015 7 Pages PDF
Abstract
In the pursuit of new natural products, the demand to rapidly identify compounds present, in ever decreasing amounts, in complex crude extracts has become a limiting factor. Despite improvements in HPLC-NMR hardware and pulse sequences, no extensive limit of detection (LOD) investigations have been reported for the acquisition of 2D NMR spectroscopic experiments acquired through HPLC-NMR. In this study the LOD for five key 1D and 2D NMR spectroscopic experiments have been established, using two reference compounds, including the on-flow (WET 1D proton), stop-flow (WET1D proton), gCOSY, HSQCAD and gHMBCAD NMR experiments. The LOD for all of the NMR experiments were within the range of 700 ng to 1 mg for the set of fixed experimental parameters implemented. For principle components in a complex multi-component mixture, this would allow for in situ compound identification. HPLC-NMR analysis was employed to investigate the principle components present in a marine brown alga crude extract, Cystophora subfarcinata.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,