Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7612983 | Journal of Chromatography A | 2014 | 26 Pages |
Abstract
A novel amphiphilic polymeric ionic liquid membrane containing a hydrophilic bromide anion and a hydrophobic carbonyl group was synthesized in dimethylformamide (DMF) systems using the ionic liquid 1-butyl-3-vinylimidazolium bromide (BVImBr) and the methylmethacrylate (MMA) as monomers. The prepared amphiphilic ploy-methylmethacrylate-1-butyl-3-vinylimidazolium bromide (MMA-BVImBr) was characterized by a scanning electron microscope and an infrared spectrum instrument. The results of solid-phase micro-extraction membrane (SPMM) experiments showed that the adsorption capacity of membrane was about 0.76 μg μgâ1 for aniline. Based on this, a sensitive method for the determination of trace aniline, as a degradation product of azo dye Orange G under sonication, was developed by liquid chromatography-tandem mass spectrometry (LC-MS/MS). The calibration curve showed a good linearity ranging from 0.5 to 10.0 μg Lâ1 with a correlation coefficient value of 0.9998. The limit of quantification was 0.5 μg Lâ1. The recoveries ranged from 90.6% to 96.1%. The intra- and inter-day relative standard deviations were less than 8.3% and 10.9%. The developed SPMM-LC-MS/MS method was used successfully for preconcentration of trace aniline produced during the sonication of Orange G solution.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Mei-Qiang Cai, Xiao-Qing Wei, Chun-Hui Du, Xu-Ming Ma, Mi-Cong Jin,