Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7613023 | Journal of Chromatography A | 2014 | 28 Pages |
Abstract
In the present work, two specific, accurate and sensitive methods for the determination of endocrine disrupting chemicals (EDCs) in human breast milk are developed and validated. Bisphenol A and its main chlorinated derivatives, five benzophenone-UV filters and four parabens were selected as target analytes. The method involves a stir-bar sorptive extraction (SBSE) procedure followed by a solvent desorption prior to GC-MS/MS or UHPLC-MS/MS analysis. A derivatization step is also necessary when GC analysis is performed. The GC column used was a capillary HP-5MS with a run time of 26 min. For UHPLC analysis, the stationary phase was a non-polar Acquity UPLC® BEH C18 column and the run time was 10 min. In both cases, the analytes were detected and quantified using a triple quadrupole mass spectrometer (QqQ). Quality parameters such as linearity, accuracy (trueness and precision), sensitivity and selectivity were examined and yielded good results. The limits of quantification (LOQs) ranged from 0.3 to 5.0 ng mLâ1 for GC and from 0.2 to 1.0 ng mLâ1 for LC. The relative standard deviation (RSD) was lower than 15% and the recoveries ranged from 92 to 114% in all cases, being slightly unfavorable the results obtained with LC. The methods were satisfactorily applied for the determination of target compounds in human milk samples from 10 randomly selected women.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
R. RodrÃguez-Gómez, A. Zafra-Gómez, F.J. Camino-Sánchez, O. Ballesteros, A. Navalón,