Article ID Journal Published Year Pages File Type
7615710 Journal of Chromatography B 2018 29 Pages PDF
Abstract
We developed and validated a new method for quantification of LTZ in human plasma and DPS. The extraction of LTZ from plasma and DPS was performed using liquid-liquid extraction with diethyl ether and an extraction solution composed of diethyl ether- methyl tert-butyl ether- acetone (50:30:20, v/v/v), respectively. Lamotrigine- 13C3, d3 was used as internal standard (ISTD) and the chromatographic separation was achieved on Hypurity Advance C18 column (150 × 4.6 mm, 5 μm). Quantitative estimation of LTZ and ISTD was performed on a liquid chromatography tandem mass spectrometer coupled with electrospray ionization interface operated under positive mode of ionization. Calibration curves were linear (r2 > 0.99) over the concentration range of 10-3020 ng/mL for both plasma and DPS. Statistical analysis provides insignificant difference between LTZ concentration extracted from plasma and DPS samples. The method is found suitable for application in clinical study and in therapeutic monitoring of LTZ. To the best of our knowledge this is the first report which describing a validated stability indicating assay for quantification of LTZ in dry plasma spot.