Abstract
Pidotimod, (R)-3-[(S)-(5-oxo-2-pyrrolidinyl) carbonyl]-thiazolidine-4-carboxylic acid, was frequently used to treat children with recurrent respiratory infections. Preclinical pharmacokinetics of pidotimod was still rarely reported to date. Herein, a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated to determine pidotimod in rat plasma, tissue homogenate and Caco-2 cells. In this process, phenacetin was chosen as the internal standard due to its similarity in chromatographic and mass spectrographic characteristics with pidotimod. The plasma calibration curves were established within the concentration range of 0.01-10.00 μg/mL, and similar linear curves were built using tissue homogenate and Caco-2 cells. The calibration curves for all biological samples showed good linearity (r > 0.99) over the concentration ranges tested. The intra- and inter-day precision (RSD, %) values were below 15% and accuracy (RE, %) was ranged from −15% to 15% at all quality control levels. For plasma, tissue homogenate and Caco-2 cells, no obvious matrix effect was found, and the average recoveries were all above 75%. Thus, the method demonstrated excellent accuracy, precision and robustness for high throughput applications, and was then successfully applied to the studies of absorption in rat plasma, distribution in rat tissues and intracellular uptake characteristics in Caco-2 cells for pidotimod.