Article ID Journal Published Year Pages File Type
7617806 Journal of Chromatography B 2014 6 Pages PDF
Abstract
A sensitive and selective ultra high performance liquid chromatography-tandem mass spectrometric (UHPLC-MS/MS) method for the determination of phlorizin and phloretin in human plasma has been firstly developed. Samples were prepared after protein precipitation and analyzed on a C18 column interfaced with a triple quadrupole tandem mass spectrometer. Negative electrospray ionization was employed as the ionization source. The mobile phase consisted of acetonitrile-water (0.02% formic acid), using a gradient procedure. The analytes and internal standard dihydroquercetin were both detected by use of multiple reaction monitoring mode. The method was linear in the concentration range of 2.5-1000.0 ng/mL. The lower limit of quantification (LLOQ) was 2.5 ng/mL. The intra- and inter-day relative standard deviation across three validation runs over the entire concentration range was less than 9.2%. The accuracy determined at three concentrations was within ±7.3% in terms of relative error. The total run time was 12.0 min. This assay offers advantages in terms of expediency, and suitability for the analysis of phlorizin and phloretin in various biological fluids.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,