| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 7617917 | Journal of Chromatography B | 2014 | 6 Pages | 
Abstract
												Electrochemical methods are very often used to detect catecholamine and indolamine neurotransmitters separated by conventional reverse-phase high performance liquid chromatography (HPLC). The present paper presents the development of a chromatographic method to detect monoamines present in low-volume brain dialysis samples using a capillary column filled with sub-2 μm particles. Several parameters (repeatability, linearity, accuracy, limit of detection) for this new ultrahigh performance liquid chromatography (UHPLC) method with electrochemical detection were examined after optimization of the analytical conditions. Noradrenaline, adrenaline, serotonin, dopamine and its metabolite 3-methoxytyramine were separated in 1 μL of injected sample volume; they were detected above concentrations of 0.5-1 nmol/L, with 2.1-9.5% accuracy and intra-assay repeatability equal to or less than 6%. The final method was applied to very low volume dialysates from rat brain containing monoamine traces. The study demonstrates that capillary UHPLC with electrochemical detection is suitable for monitoring dialysate monoamines collected at high sampling rate.
											Keywords
												
											Related Topics
												
													Physical Sciences and Engineering
													Chemistry
													Analytical Chemistry
												
											Authors
												Barbara Ferry, Elena-Patricia Gifu, Ioana Sandu, Luc Denoroy, Sandrine Parrot, 
											