Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7619156 | Journal of Chromatography B | 2012 | 8 Pages |
Abstract
⺠Fast and quantitative analysis method is presented for STX for verification purposes for CWC and also for food analyses and environmental monitoring. ⺠The developed method separates STX from other PSP toxins which are also identified by full scan MS/MS analysis. ⺠Food and Drug Administration's (FDA's) bioanalytical method validation recommendations were used and for identification, WADA criteria have been evaluated for verification of STX by using four diagnostic ions in MS/MS analysis. ⺠HILIC-MS/MS method has proven to be linear at concentration ranges from 5 to 50 ng/ml and from 25-200 ng/ml, and accurate and precise in both LLOQ and ULOQ levels. ⺠The presented HILIC-MS/MS method has been found to be applicable in proficiency testing of STX and quantitative analysis of STX in algal samples.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Mia Halme, Marja-Leena Rapinoja, Maaret Karjalainen, Paula Vanninen,