Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7619378 | Journal of Chromatography B | 2009 | 7 Pages |
Abstract
A simple and novel HPLC-MS method for the simultaneous quantification of testosterone, 11-ketotestosterone, and 11β-hydroxyandrostenedione in fish serum was developed and validated. Separation was achieved on a C-18 column using a water-acetonitrile mobile-phase with a cycle time of 12 min. Ion detection was performed using ESI positive SIM at [M+H] (m/z 303, 303, 289). The linear ranges (0.2-50 ng/ml), limits of detection (0.1-0.2 ng/ml) and quantification (0.2-0.5 ng/ml) were established. The method was validated by measuring the three androgens in goldfish sera, displaying comparable values to those reported by other analytical techniques (RIA, EIA).
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
MartÃn Blasco, Pedro Carriquiriborde, Damián Marino, Alicia E. Ronco, Gustavo M. Somoza,