Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7630477 | Journal of Pharmaceutical and Biomedical Analysis | 2014 | 7 Pages |
Abstract
A moisture quantification PLS model was developed. Twenty calibration experiments were conducted, during which spectra were collected at-line and then regressed versus the corresponding residual moisture values obtained via Karl Fischer measurements. The developed NIR moisture quantification model was then validated by calculating the accuracy profiles on the basis of the analysis results of independent in-line validation experiments. Furthermore, as the aim of the NIR method is to replace the destructive, time-consuming Karl Fischer titration, it was statistically demonstrated that the new NIR method performs at least as good as the Karl Fischer reference method.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Margot Fonteyne, Julen Arruabarrena, Jacques de Beer, Mario Hellings, Tom Van Den Kerkhof, Anneleen Burggraeve, Chris Vervaet, Jean Paul Remon, Thomas De Beer,