The optimised parameters found for the system are: flow rate 0.6 mL·minâ1, integration time 30 s and the time for pumping of solutions was 3 min, obtaining a LOD of 7.2 ppm. The CO2 response, reproducibility, precision, and stability of the sensing chemistry have been studied and compared with those obtained using bench-top instrumentation (i.e. a spectrophotometer), obtaining good agreement.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Isabel M. Perez de Vargas Sansalvador, Nuria Lopez Ruiz, M.M. Erenas, Luis Fermin Capitan Vallvey, Simon Coleman, Dermot Diamond, Maria Dolores Fernandez Ramos,