Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7643904 | Microchemical Journal | 2013 | 7 Pages |
Abstract
⺠Trace Se determination in biological samples by CH4-Ar mixed gas plasma DRC-ICP-MS ⺠The interfered 40Ar2+ on 80Se was eliminated by a dynamic reaction cell (DRC). ⺠The interfered oxide ions or double charged interfered ions were reduced up to 50%. ⺠The Se+ sensitivity was improved by a factor of 3 due to carbon-enhancement effect. ⺠Low levels of Se values of SRMs with this method agreed with the certified values.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Wei Guo, Shenghong Hu, Yanxin Wang, Luyuan Zhang, Zhaochu Hu, Jiangyi Zhang,