Article ID Journal Published Year Pages File Type
7674486 Spectrochimica Acta Part B: Atomic Spectroscopy 2014 4 Pages PDF
Abstract
A compact grazing exit X-ray fluorescence (GE-XRF) spectrometer was developed in the laboratory. An Al cylindrical collimator for the primary X-rays was placed just above the sample stage. This collimator also played the role of an exit slit to detect fluorescent X-rays at small grazing exit angles. Therefore, no additional exit slit was used in this setup, leading to a compact design. The entire size of the analysis equipment was 80 mm × 200 mm × 170 mm (horizontal × vertical × height). The maximum exit angle was adjusted to the height of the sample stage. The background was drastically reduced at grazing exit angles, enabling trace elemental analysis. A calibration curve was obtained using 10 μL Ga solutions. Accordingly, the detection limit for Ga was evaluated to be 20 ppb.