| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 7674835 | Spectrochimica Acta Part B: Atomic Spectroscopy | 2013 | 6 Pages |
Abstract
HCOOH, CH3COOH, and CH3CH2OH were used as chemical modifiers in a solution-cathode glow discharge. Emission was measured directly from the discharge, without a gas-liquid separator or a secondary excitation source. Emission from Ag, Se, Pb, and Hg was strongly enhanced, and the detection limits (DL) for these elements were improved by up to an order of magnitude using a combination of HCOOH and HNO3 compared to using HNO3 alone. The DL was measured for Mg (1 μg/L), Fe (10 μg/L), Ni (6 μg/L), Cu (6 μg/L), Pb (1 μg/L), Ag (0.1 μg/L), Se (300 μg/L), and Hg (2 μg/L). Coefficients of determination (R2) were between 0.9986 and 0.9999. A voltage of 1 kV was used, which produced a current of approximately 70 mA.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Todd A. Doroski, Michael R. Webb,
