Article ID Journal Published Year Pages File Type
7680723 Talanta 2014 36 Pages PDF
Abstract
This paper describes development and validation of a simple and efficient bioanalytical procedure for simultaneous determination of phenobarbital and carbamazepine in human serum samples using high performance liquid chromatography with photodiode-array detection (HPLC-DAD) regarding a fast elution methodology in less than 5 min. Briefly, this method consisted of a simple deproteinization step of serum samples followed by HPLC analysis on a Bonus-RP column using an isocratic mode of elution with acetonitrile/K2HPO4 (pH=7.5) buffer solution (45:55). Due to the presence of serum endogenous components as non-calibrated components in the sample, second-order calibration based on multivariate curve resolution-alternating least squares (MCR-ALS), has been applied on a set of absorbance matrices collected as a function of retention time and wavelengths. Acceptable resolution and quantification results were achieved in the presence of matrix interferences and the second-order advantage was fully exploited. The average recoveries for carbamazepine and phenobarbital were 89.7% and 86.1% and relative standard deviation values were lower than 9%. Additionally, computed elliptical joint confidence region (EJCR) confirmed the accuracy of the proposed method and indicated the absence of both constant and proportional errors in the predicted concentrations. The developed method enabled the determination of the analytes in different serum samples in the presence of overlapped profiles, while keeping experimental time and extraction steps at minimum. Finally, the serum concentration levels of carbamazepine in three time intervals were reported for morphine-dependents who had received carbamazepine for treating their neuropathic pain.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,