Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
7686295 | Talanta | 2012 | 11 Pages |
Abstract
Good linear calibrations were obtained for analytes in urine (with serotonin and dopamine giving R2 = 0.9999 and 5-HIAA having a lower R2 value of 0.9955). Acceptable intraday repeatability was achieved for all analytes with RSD values (n = 5) ranging from 4.4% to 6.2% (57, 65 and 52 nmol/L for serotonin, dopamine and 5-HIAA respectively) to 2.1-8.1% (2837, 3268, 2618 nmol/L for serotonin, dopamine and 5-HIAA respectively). Excellent limits of detection (LOD) and limits of quantitation (LOQ) were achieved with spiked samples for all compounds; with LODs of 9-12.9 nmol/L and LOQs of 27.2-57.7 nmol/L for analytes in urine. An appropriate sample clean-up procedure for urine was developed to ensure highest recovery and reproducibility on analysis.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Merisa Moriarty, Mary Lehane, Brendan O'Connell, Helen Keeley, Ambrose Furey,