Abstract
The luminiscence characteristics of thermoluminscence dosimeter LiF: Mg, Ti (TLD-100) irradiated to X-rays from 6 MV linac have been studied for wide range of 2-50 K/s readout linear heating rates. The reproducibility of glow curves for TLDs is found to be better at lower heating rates and depreciate at higher heating rates. The glow curve spectra were analysed using deconvolution procedure based on general-order kinetics. Shift in the peak maximum temperature per unit rise in heating rate for various peaks were found to decrease with heating rate. The TLDs irradiated with same dose exhibit decreasing TL counts with increase in the heating rate, which indicate the thermal quenching effect in TLD-100. The value of activation energy for each peak within the glow curve increases with heating rate. Calibration curves plotted for the dose range 0.4-1020 cGy exhibit decreasing slope with increasing readout heating rate. Corrections for temperature lag between the heating element and the dosimeter, and the effective heating rate (βeff) across the sample estimated using formulation proposed by Kitis and Tuyn and are found to be fairly applicable.