Abstract
However, the technical ability to identify 36Cl ions is quite distinct from demonstrated high-performance AMS. Such is the theme of this paper. We present a systematic analysis of the accurate measurement of sample radioisotope relative to the stable chlorine, the normalisation of the measured ratio and correction for remaining 36S interference, all combined with the use of stable-isotope dilution to determine sample Cl concentration to begin with. We conclude by showing that repeated analyses support our claims for routine 3% 36Cl-AMS data. Accordingly, the modest SUERC spectrometer well competes with the performance of larger longer-established instruments, and the results may be quite generic for modern bespoke instruments.