Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8081839 | Journal of Environmental Radioactivity | 2016 | 5 Pages |
Abstract
The aim of this paper was to calculate the True Coincidence Summing Correction Factors (TSCFs) for an HPGe coaxial detector in order to correct the summing effect as a result of the presence of 88Y and 60Co in a multigamma source used to obtain a calibration efficiency curve. Results were obtained for three volumetric sources using the Monte Carlo toolkit, GEANT4. The first part of this paper deals with modeling the detector in order to obtain a simulated full energy peak efficiency curve. A quantitative comparison between the measured and simulated values was made across the entire energy range under study. The True Summing Correction Factors were calculated for 88Y and 60Co using the full peak efficiencies obtained with GEANT4. This methodology was subsequently applied to 134Cs, and presented a complex decay scheme.
Related Topics
Physical Sciences and Engineering
Energy
Nuclear Energy and Engineering
Authors
G. Giubrone, J. Ortiz, S. Gallardo, S. Martorell, M.C. Bas,