Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8131501 | Ultrasound in Medicine & Biology | 2017 | 5 Pages |
Abstract
The aim of this study was to perform a blinded trial, comparing the results of a visual inspection of the in-air reverberation pattern with the results of an electronic probe tester in detecting ultrasound probe faults. Sixty-two probes were tested. A total of 28 faults were found, 3 only by in-air reverberation assessment and 2 only by the electronic probe tester. The electronic probe tester provided additional information regarding the location of the fault in 74% of the cases in which both methods detected a fault. It is possible to detect the majority of probe faults by visual inspection and in-air reverberation assessment. The latter provides an excellent first-line test, easily performed on a daily basis by equipment users. An electronic probe tester is required if detailed evaluation of faults is necessary.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Acoustics and Ultrasonics
Authors
Nicholas J. Dudley, Darren J. Woolley,