Article ID Journal Published Year Pages File Type
8147090 Infrared Physics & Technology 2015 5 Pages PDF
Abstract
How to select optimal ADC parameters in indirect measurement including resolution, sampling rate and sampling error of ADC, and system error is proposed. Fitting of the photo time constant in cavity ring-down spectrum is taken as an example. The effects of resolution, sampling rate and sampling error (with/without system error) of ADC on fitting accuracy are simulated sequentially. The influence of mirror reflectivity on the parameters of fitting functions is analyzed and simulated. Parameters of ADC to distinguish concentration of methane of 1 ppbv at different mirror reflectivity are given.
Keywords
Related Topics
Physical Sciences and Engineering Physics and Astronomy Atomic and Molecular Physics, and Optics
Authors
, , , ,