Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8166284 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2018 | 17 Pages |
Abstract
The neutron field measurement was performed in the Experimental Advanced Superconducting Tokamak (EAST) experimental hall using a Bonner sphere spectrometer (BSS) based on a 3He thermal neutron counter. The measured spectra and the corresponding integrated neutron fluence and dose values deduced from the spectra at two exposed positions were compared to the calculated results obtained by a general Monte Carlo code MCNP5, and good agreements were found. The applicability of a homemade dose survey meter installed at EAST was also verified with the comparison of the ambient dose equivalent H*(10) values measured by the meter and BSS.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Zhimeng Hu, Guoqiang Zhong, Lijian Ge, Tengfei Du, Xingyu Peng, Zhongjing Chen, Xufei Xie, Xi Yuan, Yimo Zhang, Jiaqi Sun, Tieshuan Fan, Ruijie Zhou, Min Xiao, Kai Li, Liqun Hu, Jun Chen, Hui Zhang, Giuseppe Gorini, Guohui Zhang,