Article ID Journal Published Year Pages File Type
8166296 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2018 7 Pages PDF
Abstract
This paper describes a method for more complete neutron detector characterization using Geant4's Monte Carlo methods for characterizing overall detector response rate and Garfield interfaced with SRIM for the simulation of the detector's raw pulses, as applied to a hybrid 10B/3He detector. The Geant4 models characterizing the detector's interaction with a 252Cf point source and parallel beams of mono-energetic neutrons (assuming ISO 8529 reference energy values) compare and agree well with calibrated 252Cf measurements to within 6.4%. Validated Geant4 model outputs serve as input to Garfield+SRIM calculations to provide meaningful pulse height spectra. Modifications to Garfield for this work were necessary to account for simultaneous tracking of electrons resulting from proton and triton reaction products from a single 3He neutron capture event, and it was further necessary to interface Garfield with the energy loss, range, and straggling calculations provided by SRIM. Individual raw pulses generated by Garfield+SRIM are also observed to agree well with experimentally measured raw pulses from the detector.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,