Abstract
In this work, we characterize the γ-ray response and efficiency for a cylindrical inorganic Cs2LiYCl6:Ce detector 1′′ in diameter and 1′′ in height. The energy resolution and linearity are obtained from 21 γ-rays with energies ranging from 0.026 to 2.447 MeV. In addition, the neutron γ-ray discrimination is validated by measuring a 252Cf radioisotope. Gamma-ray response functions and matrix below 7 MeV are simulated using a Monte Carlo approach and validated through the unfolded γ-ray spectra.