We have developed a beam-profile monitor (BPM) system to align the collimators for the neutral beam-line at the Hadron Experimental Facility of J-PARC. The system is composed of a phosphor screen and a CCD camera coupled to an image intensifier mounted on a remote control X-Y
stage. The design and detailed performance studies of the BPM are presented. The monitor has a spatial resolution of better than 0.6Â mm and a deviation from linearity of less than 1%. These results indicate that the BPM system meets the requirements to define collimator-edge positions for the beam-line tuning. Confirmation using the neutral beam for the KOTO experiment is also presented.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
T. Matsumura, I. Kamiji, K. Nakagiri, H. Nanjo, T. Nomura, N. Sasao, T. Shinkawa, K. Shiomi,