Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8166993 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2018 | 7 Pages |
Abstract
The paper presents a study of nâγ discrimination with 4x4 ch and 8x8 ch Multi Pixel Photon Counter (MPPC) arrays in neutron detectors based on Stilbene and EJ299-33 plastic scintillators. The nâγ discrimination showed an excellent capability of the MPPC arrays, comparable to that observed earlier with the classical PMTs. Particularly, an application of a zero-crossing method of nâγ discrimination prevented deterioration of the discrimination by the slow response of the Silicon Photomultiplier (SiPM, or MPPC interchangeably) array related to its large capacitance. It was confirmed by a good agreement of the Figure of Merit normalized to the number of photoelectrons determined for the MPPC arrays and XP5500 PMT.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
M. Grodzicka-Kobylka, T. Szczesniak, M. MoszyÅski, L. Swiderski, D. Wolski, J. Baszak, S. Korolczuk, P. Schotanus,