Article ID Journal Published Year Pages File Type
8167165 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2018 6 Pages PDF
Abstract
In this work we present the 2 MV Tandetron accelerator manufactured by High Voltage Engineering Europa (HVEE), that was installed at MTA Atomki in Debrecen, Hungary in 2015. Furthermore, we report on the first performance test of the whole facility including the calibration of the terminal voltage using accurately known resonance energies of the  27Al(p,γ)28Si and 13C(p,γ)14N reactions, as well as neutron threshold energies of the  7Li(p,n)7Be and  13C(p,n)13N reactions. The accurate energy calibration of the generating voltmeter (GVM), the good energy stability of the accelerator as well as the low energy spread of the ion beam suited the requirements of nuclear astrophysics, resonance PIGE depth profiling and the nanoprobe beamline. We have investigated whether the terminal voltage reading depends on the insulating gas pressure, and how much the stripper gas pressure modifies the beam energy.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , , ,