8167198 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2018 | 13 Pages |
Abstract
The aim of this paper is to characterize two HPGe gamma-ray detectors used in two different laboratories for environmental radioactivity measurements, so as to perform efficiency calibrations by means of Monte Carlo Simulation. To achieve such an aim, methodologies developed in previous papers have been applied, based on the automatic optimization of the model of detector, so that the differences between computational and reference FEPEs are minimized. In this work, such reference FEPEs have been obtained experimentally from several measurements of the IAEA RGU-1 reference material for specific source-detector arrangements. The models of both detectors built through these methodologies have been validated by comparing with experimental results for several reference materials and different measurement geometries, showing deviations below 10% in most cases.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
J. G. Guerra, J. G. Rubiano, G. Winter, A. G. Guerra, H. Alonso, M.A. Arnedo, A. Tejera, F. Mosqueda, P. Martel, J.P. Bolivar,