Article ID Journal Published Year Pages File Type
8167947 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2017 4 Pages PDF
Abstract
Tumour uptake value is a critical result in [18F]FDG-PET/CT ([18F]fluorodeoxyglucose) quantitative scans such as the dose prescription for radiotherapy and oncology. The quantification is highly dependent on the protocol of acquisition and reconstruction of the image, especially in low activity tumours. During adjusting acquisition and reconstruction protocols available in our Siemens Biograph mCT scanner for EARL (ResEARch 4 Life®) [18F]FDG-PET/CT accreditation requirements, we developed reconstruction protocols which will be used in PET based radiotherapy planning able to reduce inter-/intra-institute variability in Standard Uptake Value (SUV) results, and to bring Recovery Coefficient to 1 as close as possible for Image Quality NEMA 2007 phantom. Primary and secondary tumours from two patients were assessed by four independent evaluators. The influence of reconstruction protocols on tumour clinical assessment was presented. We proposed the improvement route for EARL accredited protocols so that they may be developed in classes to take advantage of scanner possibilities. The application of optimized reconstruction protocol eliminates the need of partial volume corrections.
Keywords
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , , , , , ,