A lightweight, low-cost multi-panel direction-sensitive radiation detector prototype has been developed at Oregon State University that is designed to be mounted on a small unmanned aerial system to autonomously search for radiation sources while flying close to the ground. The detection system comprises sixteen BGO-SiPM detector panels with an adjustable view angle, and signal outputs are processed in parallel in an FPGA. The minimum detectable activity was calculated to be 1.3 μCi of 137Cs at 1 m in under 60 s. The counting response of the detector panels were characterized and found to have 4.7% relative standard deviation, indicating good uniformity in overall design and assembly. The detector was also able to estimate the direction of a 12.3 μCi 137Cs source 100 cm from the device center with 2.3° accuracy in a 95% confidence width of 10.8° in 60 s.