Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8168219 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2016 | 5 Pages |
Abstract
We report the measurement of the emission time profile of scintillation from gamma-ray induced events in the XMASS-I 832Â kg liquid xenon scintillation detector. Decay time constant was derived from a comparison of scintillation photon timing distributions between the observed data and simulated samples in order to take into account optical processes such as absorption and scattering in liquid xenon. Calibration data of radioactive sources, 55Fe, 241Am, and 57Co were used to obtain the decay time constant. Assuming two decay components, Ï1 and Ï2, the decay time constant Ï2 increased from 27.9Â ns to 37.0Â ns as the gamma-ray energy increased from 5.9Â keV to 122Â keV. The accuracy of the measurement was better than 1.5Â ns at all energy levels. A fast decay component with Ï1â¼2ns was necessary to reproduce data. Energy dependencies of Ï2 and the fraction of the fast decay component were studied as a function of the kinetic energy of electrons induced by gamma-rays. The obtained data almost reproduced previously reported results and extended them to the lower energy region relevant to direct dark matter searches.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation