Article ID Journal Published Year Pages File Type
8169953 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2016 4 Pages PDF
Abstract
Quality assurance in hadrontherapy remains an open issue that can be addressed with reliable monitoring of treatment accuracy. The INSIDE (INnovative SolutIons for DosimEtry in hadrontherapy) project aims to develop an integrated online monitoring system based on two dedicated PET panels and a tracking system, called Dose Profiler. The proposed solution is designed to operate in-beam and provide an immediate feedback on the particle range acquiring both photons produced by β+ decays and prompt secondary particle signals. Monte Carlo simulations cover an important role both in the system development, by confirming the design feasibility, and in the system operation, by understanding data. A FLUKA-based integrated simulation was developed taking into account the hadron beam structure, the phantom/patient features and the PET detector and Dose Profiler specifications. In addition, to reduce simulation time in signal generation on PET detectors, a two-step technique has been implemented and validated. The first PET modules were tested in May 2015 at the Centro Nazionale Adroterapia Oncologica (CNAO) in Pavia (Italy) with very satisfactory results: in-spill, inter-spill and post-treatment PET images were reconstructed and a quantitative agreement between data and simulation was found.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation