Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
8170164 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2016 | 7 Pages |
Abstract
The excess noise factor and the aging characteristics of 16 Hamamatsu R11283 photopentodes have been tested. These fine-mesh phototubes are to be paired with pure CsI scintillation crystals considered for use in the endcap calorimeter of the Belle II detector. The average excess noise factor was found to be 1.9±0.1±0.4. The electronic noise of a custom preamplifier produced by the University of Montreal was found as a consequence of this measurement and was 1730±33 electrons, in agreement with previous values. On average, the gainÃquantum efficiency was reduced to 92±3% of the initial value after passing an average of 7 C through the anode. This corresponds to 70 years of standard Belle II operation.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
D. Fujimoto, C. Hearty,